Define B3 sensor for calibration
For B3 automatic calibration a dialog is opened to select the Boomerang 3 sensor that will be calibrated.

You may search for a specific name using the field on top of the dialog.

Tap the sensor you want to calibrate. A form is opened to enter the serial number and resolution of the sensor to calibrate. The form contains also the environment conditions.
The serial number is stated on the sensor label. Some of the fields may be filled in automatically, basing on B3 settings available for a given MP.
In "Calibrated sensor resolution" select the correct sensor type or select "Custom" and enter the resolution manually. The detailed information about sensor resolutions can be found in Calibrated sensor resolution.
Enter the delay (min. 4 min) and the total time (min. 15 min).

Tap “OK” to start the calibration. The app will communicate with Boomerang 3 server to initiate the calibration process.
In case of WLREF sensor, press its button to wake it up as the Cal App must send a command to initiate the measurements. After initiation, WLREF will start measuring every minute and will store the readings in its internal memory.

Next follow the steps described in Calibration run.