Reference readings adjustment
The WLREF reference readings are adjusted basing on the data stored in WLREF adjustment file.
Basing on the data stored in the adjustment file, an adjustment is defined for every refererence reading. The app selects the adjustment for the calibration temperature which is the closest to the reference raw reading.
The raw reference reading is then adjusted and rounded to 2 digits after decimal point.
Example
WLREF adjustments file
ceb3451dfbc9
246/2/25
-80,005 0,023 0,030
-55,003 0,011 0,030
-40,000 0,014 0,030
-20,003 0,020 0,030
-0,002 0,018 0,030
4,999 -0,006 0,030
9,999 -0,004 0,030
20,000 0,006 0,030
30,005 0,017 0,030
37,004 0,012 0,030
49,998 -0,003 0,030
105,005 0,013 0,030The reference readings, their adjustments, adjusted values and adjusted rounded values. Reading 1 and 2 have adjustment values taken from 20°C row, all the other from 105°C row.
Reading 1
24.414
0.006
24.408
24.41
Reading 2
24.273
0.006
24.267
24.27
Reading 3
89.218
0.013
89.205
89.21
Reading 4
87.000
0.013
86.987
86.99
Reading 5
85.500
0.013
85.487
85.49
Reading 6
84.187
0.013
84.174
84.17
Reading 7
83.109
0.013
83.096
83.10
Reading 8
82.093
0.013
82.080
82.08
Reading 9
81.132
0.013
81.119
81.12
Reading 10
80.242
0.013
80.229
80.23
Last updated