The WLREF reference readings are adjusted basing on the data stored in WLREF adjustment file.

Basing on the data stored in the adjustment file, an adjustment is defined for every refererence reading. The app selects the adjustment for the calibration temperature which is the closest to the reference raw reading.

The raw reference reading is then adjusted and rounded to 2 digits after decimal point.

Example

WLREF adjustments file

ceb3451dfbc9
246/2/25
-80,005	0,023	0,030
-55,003	0,011	0,030
-40,000	0,014	0,030
-20,003	0,020	0,030
-0,002	0,018	0,030
4,999	-0,006	0,030
9,999	-0,004	0,030
20,000	0,006	0,030
30,005	0,017	0,030
37,004	0,012	0,030
49,998	-0,003	0,030
105,005	0,013	0,030

The reference readings, their adjustments, adjusted values and adjusted rounded values. Reading 1 and 2 have adjustment values taken from 20°C row, all the other from 105°C row.

Reference series readings
Raw
Adjustment
Adjusted
Adjusted rounded

Reading 1

24.414

0.006

24.408

24.41

Reading 2

24.273

0.006

24.267

24.27

Reading 3

89.218

0.013

89.205

89.21

Reading 4

87.000

0.013

86.987

86.99

Reading 5

85.500

0.013

85.487

85.49

Reading 6

84.187

0.013

84.174

84.17

Reading 7

83.109

0.013

83.096

83.10

Reading 8

82.093

0.013

82.080

82.08

Reading 9

81.132

0.013

81.119

81.12

Reading 10

80.242

0.013

80.229

80.23

Last updated