# Alarm port

<figure><img src="/files/UOkp2UeBXtDhlvUCJKSe" alt=""><figcaption></figcaption></figure>

<figure><img src="/files/huNZxFrpWTdgf30gltl8" alt="" width="212"><figcaption></figcaption></figure>

The external alarm port is an isolated switch output with the maximum current rating of 1.0 A and maximum voltage of 50V AC or DC.&#x20;

The switch is closed when there is no alarm and opens when alarm condition is detected.

The switch is connected to pins 1 and 2 of the XLR socket.

Note that there is an error in the incubator manual where the picture points to pins 1 and 2 but text tells to use pins 1 and 3. The text is incorrect and shall be corrected by the producer.

<figure><img src="/files/IBRcRpLlEUnG0vtsjzh9" alt=""><figcaption></figcaption></figure>

Solder two FK002 cables to pins 1 and 2 of the XLR female plug.

## B3 settings

In B3, FK002 sensor is used to observe the alarm status. Solder the FK002 cables to pins 1 and 2 of the XLR plug.

The alarm shall be raised when an opened state is detected.


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://help.boomerang3.com/external-sensors-in-incubators/ivftech-freygen/alarm-port.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.